OpenPolScope MantisBT

View Issue Details Jump to Notes ] [  ] Issue History ] Print ]
IDProjectCategoryView StatusDate SubmittedLast Update
0000085OpenPolScope-API[All Projects] bugpublic2018-08-06 18:042018-08-06 18:07
Reporterrudolfo 
Assigned Toamit 
PriorityhighSeveritymajorReproducibilityalways
StatusassignedResolutionopen 
PlatformNikon Ti stationOSOS Version
Product Version 
Target VersionFixed in Version 
Summary0000085: wavelength change not complete in birefringence mode
DescriptionAfter taking retardance images with one wavelength, I changed the wavelength on the scope and entered the new wavelength into the MeadowlarkLC-Controls panel of the Pol-Acquisition window. I then calibrated with the new wavelength, which finished in time, but showed settings for LC-A and LC-B that were quite far off from the usual 0.25 and 0.5. After using the Device Property Browser tool, it calibrated correctly.
My hunch is that the wavelength is initially not set correctly for the first calibration. Might it first use the default value 546nm?
TagsNo tags attached.
Attached Files

- Relationships

-  Notes
There are no notes attached to this issue.

- Issue History
Date Modified Username Field Change
2018-08-06 18:04 rudolfo New Issue
2018-08-06 18:07 rudolfo Assigned To => amit
2018-08-06 18:07 rudolfo Status new => assigned