View Issue Details
0000085OpenPolScope-API[All Projects] bugpublic2018-08-06 18:042018-08-06 18:07
rudolfo 
amit 
highmajoralways
assignedopen 
Nikon Ti station
 
 
0000085: wavelength change not complete in birefringence mode
After taking retardance images with one wavelength, I changed the wavelength on the scope and entered the new wavelength into the MeadowlarkLC-Controls panel of the Pol-Acquisition window. I then calibrated with the new wavelength, which finished in time, but showed settings for LC-A and LC-B that were quite far off from the usual 0.25 and 0.5. After using the Device Property Browser tool, it calibrated correctly.
My hunch is that the wavelength is initially not set correctly for the first calibration. Might it first use the default value 546nm?
No tags attached.
Issue History
2018-08-06 18:04rudolfoNew Issue
2018-08-06 18:07rudolfoAssigned To => amit
2018-08-06 18:07rudolfoStatusnew => assigned

There are no notes attached to this issue.