OpenPolScope MantisBT - OpenPolScope-API | |||||
| View Issue Details | |||||
| ID | Project | Category | View Status | Date Submitted | Last Update |
| 0000085 | OpenPolScope-API | [All Projects] bug | public | 2018-08-06 18:04 | 2018-08-06 18:07 |
| Reporter | rudolfo | ||||
| Assigned To | amit | ||||
| Priority | high | Severity | major | Reproducibility | always |
| Status | assigned | Resolution | open | ||
| Platform | Nikon Ti station | OS | OS Version | ||
| Product Version | |||||
| Target Version | Fixed in Version | ||||
| Summary | 0000085: wavelength change not complete in birefringence mode | ||||
| Description | After taking retardance images with one wavelength, I changed the wavelength on the scope and entered the new wavelength into the MeadowlarkLC-Controls panel of the Pol-Acquisition window. I then calibrated with the new wavelength, which finished in time, but showed settings for LC-A and LC-B that were quite far off from the usual 0.25 and 0.5. After using the Device Property Browser tool, it calibrated correctly. My hunch is that the wavelength is initially not set correctly for the first calibration. Might it first use the default value 546nm? | ||||
| Steps To Reproduce | |||||
| Additional Information | |||||
| Tags | No tags attached. | ||||
| Relationships | |||||
| Attached Files | |||||
| Issue History | |||||
| Date Modified | Username | Field | Change | ||
| 2018-08-06 18:04 | rudolfo | New Issue | |||
| 2018-08-06 18:07 | rudolfo | Assigned To | => amit | ||
| 2018-08-06 18:07 | rudolfo | Status | new => assigned | ||
| There are no notes attached to this issue. |